• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE MOT DR UNIT HNDCONT POWERMAX ELITE; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE MOT DR UNIT HNDCONT POWERMAX ELITE; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200872S
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that the motor drive unit had an unspecified functional failure.Incident occurred during set up/inspection of the device; therefore, there was no patient involvement.Results of investigation have concluded that this unit had a motor blocked which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issues were observed.A functional evaluation found a blocked motor.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE MOT DR UNIT HNDCONT POWERMAX ELITE
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11016986
Report Number1643264-2020-02108
Device Sequence Number10951036
Product Code HAB
UDI-Device Identifier03596010653703
UDI-Public3596010653703
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72200872S
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/07/2020
Supplement Date Received by Manufacturer02/05/2021
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date03/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-