| Model Number |
H7493941832500 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first date of the month of the aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 5.00 x 32mm synergy xd drug-eluting stent was advanced for treatment.However, during procedure, it was noticed that there was a blood came back up into the endoflator which was attached to the monorail portion of the stent.When the device was pulled out, a hole in the balloon portion of the stent delivery system was noted.The procedure was completed with another stent and no patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 5.00 x 32mm synergy xd drug-eluting stent was advanced for treatment.However, during procedure, it was noticed that there was a blood came back up into the endoflator which was attached to the monorail portion of the stent.When the device was pulled out, a hole in the balloon portion of the stent delivery system was noted.The procedure was completed with another stent and no patient complications were reported.It was further reported that the 80%-90% stenosed, non-tortuous, eccentric target lesion was located in the mid right coronary artery.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event was corrected from (b)(6) 2020.
|
| |
|