| Model Number |
357466 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Two passeo-18 balloon catheters were chosen for the treatment during the same procedure.The balloons burst during procedure.The second passeo-18 is reported separately.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint instrument was not provided for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations no manufacturing related root cause could be determined.
|
| |
|