• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/18; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/18; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364491
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a moderately calcified lesion (75 percent stenosis degree) in a moderately tortuous rca.The device could no pass the lesion.Another orsiro was used to proceed with the intervention.
 
Additional Manufacturer Narrative
Combination product: yes the returned device was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon is well folded and shows no signs of inflation.The stent does not show any damage or irregularity.The crimped diameter of the stent still complies with the specification.Review of the production documentation of the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11018723
Report Number1028232-2020-05510
Device Sequence Number14351090
Product Code NIQ
UDI-Device Identifier07640130439061
UDI-Public7640130439061
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model Number364491
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03194788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Initial Date Received by Manufacturer 12/15/2020
Supplement Date Received by Manufacturer02/23/2021
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date02/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-