• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 425-00
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The complaint is reported as: there is an issue with the temperature probe causing the unit to alarm.The issue was detected prior to use on a patient during pre-testing.No patient involvement.
 
Event Description
The complaint is reported as: there is an issue with the temperature probe causing the unit to alarm.The issue was detected prior to use on a patient during pre-testing.No patient involvement.
 
Manufacturer Narrative
Qn# (b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.No abnormal alarms were observed.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key11018864
MDR Text Key221770891
Report Number3003898360-2020-00982
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Initial Date Manufacturer Received 11/23/2020
Initial Date FDA Received12/16/2020
Supplement Dates Manufacturer Received02/05/2021
Supplement Dates FDA Received02/05/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-