| Model Number |
405116 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Insert (1316)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Pneumothorax (2012)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
During an implantable cardioverter-defibrillator procedure, the sheath peeled at the distal end while being inserted.Following the procedure, a pneumothorax and a compartment hematoma were noted.
|
| |
|
Event or Problem Description
|
|
During an implantable cardioverter-defibrillator procedure, the sheath peeled at the distal end while being inserted.The sheath was replaced and the procedure was completed.Following the procedure, a pneumothorax was diagnosed via x-ray and a compartment hematoma were noted.The vascular puncture was noted to be closed normally.The patient was sent to intensive care for closer monitoring.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8f peel-away introducer sheath was received for evaluation.The dilator was also received.The reported event of the ¿sheath peeled at the distal end while being inserted¿ was confirmed.The sheath distal tip had been split along both scorelines; the sheath distal tip had been split and there were multiple bends throughout the sheath tubing distal end.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported hematoma and pneumothorax remains unknown.The cause of the ¿sheath peeled at the distal end while being inserted¿ is consistent with damage during use.
|
| |
|