BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Codes Material Rupture (1546); Material Deformation (2976)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon tear occurred.A 3.0x40x90 sterling balloon catheter was selected for use.However, when the balloon was taken out of the sterile packet, it was misshaped.The proximal part of the balloon was bent and part of the balloon was split opened.The procedure was completed with another of the same device.No patient complications were reported.
 
Event or Problem Description
It was reported that balloon rupture occurred.A 3.0x40x90 (4f) sterling balloon catheter was selected for use.However, when the balloon was taken out of the sterile packet and it was misshaped.He said the proximal part of the balloon was bent and part of the balloon was split open.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling otw balloon catheter.Analysis of the tip, balloon and shaft included microscopic and visual inspection.Inspection revealed that the balloon was tightly folded until 43mm (from tip) and then the balloon was observed to be open/damaged.Inspection of the rest of the device found no other damage or defect.The reported balloon damage and bend were confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11019518
Report Number2134265-2020-17936
Device Sequence Number3823451
Product Code LIT
UDI-Device Identifier08714729860143
UDI-Public08714729860143
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K141112
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/11/2021
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0025515749
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2021
Initial Date Received by Manufacturer 12/09/2020
Supplement Date Received by Manufacturer01/26/2021
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-