DEPUY ORTHOPAEDICS INC US PINN 100 W/GRIPTION 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-31-054
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon put pinnacle cup (pinnacle 100 gription size 54mm) to the patient and tried to put apex hole eliminator to place, but it came through the hole.He took a new apex hole eliminator and same happened.He took the cup away and put a new cup without apex hole eliminator.Surgery delayed by 10-15min.The patient did not suffer any harm and has recovered well.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A search of the depuy nonconformance (nc) quality system found no manufacturing related nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a search of the depuy nonconformance (nc) quality system found no manufacturing related nc¿s associated with this product/lot combination.Device history review : a search of the depuy nonconformance (nc) quality system found no manufacturing related nc¿s associated with this product/lot combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN 100 W/GRIPTION 54MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11019661
Report Number1818910-2020-27045
Device Sequence Number19819471
Product Code JDI
UDI-Device Identifier10603295010210
UDI-Public10603295010210
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K090998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-31-054
Device Catalogue Number121731054
Device Lot Number9512738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 12/02/2020
Supplement Date Received by Manufacturer12/02/2020
12/15/2020
01/14/2021
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date12/21/2020
12/23/2020
01/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; APEX HOLE ELIM POSITIVE STOP
Patient Age55 YR
-
-