| Catalog Number |
512535VER |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a diagnostic neuro angiogram procedure, the catheter tip detached within the patient.The physician used a vascular snare device to successfully remove the foreign body from the patient with no additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.A photo attached to this case verifies that the catheter tip has separated.Since the tip is separated in the photo the complaint is confirmed.Review of attached photo showed two small elongated sections of the tip on both sides of the fracture.The root cause could not be determined however, it is likely that significant force was applied to the device during clinical use.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned at a later date, the investigation will be re-opened.
|
| |
|