| Catalog Number |
814501095 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Fall (1848); Pain (1994); Osteopenia/ Osteoporosis (2651)
|
| Date of Event |
10/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-03729.Concomitant medical devices: part# 814511180; lot# unk; part# 814510115; lot# vk1121315a; part# 814550040; lot# unk.Foreign - event occurred in denmark.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported approximately one (1) year and four (4) months ago, patient was implanted with an unknown affixus nail to repair a trochanter femoris fracture on the right side.Approximately seven (7) months post-op, the patient fell.Xrays were taken approximately four (4) months ago due to patient experiencing pain, and x-rays had shown the nail had fractured.Patient was revised approximately two (2) months ago to remove the nail.During the surgery, a screw was also discovered to have fractured.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned products noted the nail and screw fracture.The returned parts were sent to sem for additional evaluation.For screw - suspected fatigue crack initiation area near the edge of the screw, which appeared to be smeared.Fracture surface revealed artifacts that were consistent with fatigue mode, exhibiting fatigue striations.Suspected crack exit area identified near the opposite end of the screw fracture showing the transition from fatigue to overload mode of fracture with ductile dimples.Medical records were not provided.X-rays provided were sent to an external hcp for review.Their review noted the following: implant fit and alignment appear maintained although as noted one of the femoral neck surgical screws is fractured.Bone quality is osteopenic.Device history records was reviewed and no discrepancies relevant to the reported event were found.It was mentioned that the patient fell in (b)(6) 2020, which could potentially lead to the fracture of the implants.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|