• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH QC F/DHS/DCS TRIPLE REAMERS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH QC F/DHS/DCS TRIPLE REAMERS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.015
Medical Device Problem Codes Device Slipped (1584); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event or Problem Description
It was reported from (b)(6) that the base of the triple reamer attachment device could be unscrewed and the gear was loose.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had a worn etch, coupling could not be engaged, illegible etch, coupling was loose, component damage and failed pre-test for tool quick coupling, machine coupling, marking and labeling and general condition.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QC F/DHS/DCS TRIPLE REAMERS
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11023168
Report Number8030965-2020-09744
Device Sequence Number19957246
Product Code HWE
UDI-Device Identifier07611819107168
UDI-Public07611819107168
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number532.015
Device Lot NumberMA1079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2020
Supplement Date Received by Manufacturer12/23/2020
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-