It was reported that the trocars on the bard drains were dull and dangerous.In addition, the same issue was found with the same drains used at (b)(6) facility.Per follow up via email on (b)(6) 2020, as trocar has been inserted with extreme difficulty and patient needed restitching of fascia that came apart while attempting to insert trocar.All issue occurred during surgical procedure.No extra intervention required apart from extra stitches.
|
The reported event was inconclusive as no sample returned for evaluation.A potential root cause for this failure mode could be due to the trocar tip was not sharp enough to create the exit site which resulted in damage tip or tip did not meet the specifications.It was unknown whether the device had met specifications.The product used for the treatment but it was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number was unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications: wound drains are used to remove exudates from wound sites.Warning: when placing drain(s) care should be taken to ensure that the perforated portion of the wound drain lies completely within the confines of the wound.Read product insert provided with the closed wound evacuator for detailed instructions, warnings and precautions associated with the use of the evacuator device.To avoid the possibility of drain damage or breakage: additional perforations should not be made in the drains.Avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Drain placement: 1.Place perforated wound drain within the critical fluid collection area of wound.2.Draw non-perforated section of wound drain through to the outside until drain indicator mark appears at the skin surface.Two sets of indicator marks aid placement of the drain.3.Remove trocar only by cutting the drain one inch from end of trocar.4.Trim non-perforated section of drain to desired length.For bard 100cc silicone and reliavac 100 evacuators.Single drain placement: 1.Remove adapter with connector from pouch.2.Insert adapter end into non-perforated section of drain.3.Insert connecting tube of drain to adapter port of evacuator.For two drain placement: for bard 100cc silicone evacuators.1.Attach second drain to adapter.2.Attach connecting tube of both silicone drains to the bard® y-connector and connect y-connector to evacuator drain port.For reliavac 100 (use blue adapter in drain package and y-connector in evacuator package).1.Cut off plug from closed arm of y-connector.2.Remove blue adapters from pouches.Insert into y-connector.3.Attach both drains to blue adapters.For reliavac 400 and 3-spring evacuators.Single drain placement: 1.Ensure open end of y-connector is open or cut to one fourth mark.2.Remove blue adapter from pouch.Insert into y-connector.3.Attach drain to blue adapter.For two drain placement: 1.Cut off plug from closed arm of y-connector.2.Remove blue adapter from pouch.Insert into y-connector.3.Attach both drains to blue adapters.Reuse precaution: this is a single use device.Do not resterilize any portion of this device.Reuse and or repackaging may create a risk of patient or user infection, compromise the structural integrity and or essential material and design characteristics of the device, which may lead to device failure, and or lead to injury, illness or death of the patient." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|