| Model Number |
2911-05545 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Devices remains implanted.
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-operative final tightening of an everest polyaxial screw at t5, the screw tulip 'splayed'.The screw was left inside the patient as is.There were no adverse consequences to the patient.The procedure was completed successfully with a 1-minute delay to surgery by leaving the screw in place.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, functional and material analysis could not be performed as the device was not returned.A review of device and complaint history records could not be performed as a valid lot code was not provided and could not be obtained.Per email correspondence, tulip was not over angulated.Final tightening was performed using torque wrench and anti torque handle, excessive force was not applied.Due to device not returned, an exact cause of the reported event could not be determined.Causes related to this event include: excess torque applied while final tightening, torque wrench not calibrated, cantilever force applied, etc.
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-operative final tightening of an everest polyaxial screw at t5, the screw tulip 'splayed'.The screw was left inside the patient as is.There were no adverse consequences to the patient.The procedure was completed successfully with a 1-minute delay to surgery by leaving the screw in place.
|
| |
|