| Model Number |
TV-IL1414160-J |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
Occlusion (1984)
|
| Date of Event |
12/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with an ovation ix stent graft system to treat an abdominal aortic aneurysm (aaa).Approximately two (2) months post initial procedure there was an occluded left limb and cross limbs were identified.A secondary procedure was performed.The physician performed an open thrombectomy then placed stents and an ovation iliac limb extension to cover the clot.The patient was reported as doing fine post secondary procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the left limb stent occlusion and secondary endovascular procedure events are confirmed.This is consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported stable.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g1: contact office; g4: awareness date; h6: result code: remove code 3233; h6: conclusion code: remove code 11.
|
| |
|
Search Alerts/Recalls
|