Model Number 0682400550
Medical Device Problem Code Electromagnetic Interference (1194)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that an equipment boom electrical rotational brakes disengaged when a cauterizer device was being used during a procedure.There was patient involvement due to the issue occurring during a procedure but there was no impact to the patient and caused no surgical delay.They experienced the issue when the patient was closed up.There were no reported injuries or adverse consequences.A stryker field service technician (sfst) was dispatched to the account to perform an investigation.The sfst confirmed the failure with the customer but was unable to recreate the issue while on-site.The sfst observed that the equipment boom mfr assembly had the old-style revision of capacitive touch board although we couldn't confirm the root cause we suspect the issue is most likely due to the previous version of mfr capacitive board.The newer revision capacitive touch board was redesigned and released back in april 2020 to be less susceptible to emi from high frequency devices.The sfst was unable to install the newer mfr capacitive touch board due to parts being backordered, once parts are available we will install the redesigned capacitive touch board.If any further information around this event is obtained we will file a supplemental.
 
Event or Problem Description
It was reported the cautery was in use during a case and the boom brakes disengaged in or 6.There were no reported injuries or adverse consequences.
 
Additional Manufacturer Narrative
It was reported that once parts were available, the stryker field service technician (sfst) installed our current released mfr board and confirmed that the boom was operational.
 
Event or Problem Description
It was reported the cautery was in use during a case and the boom brakes disengaged in or 6.There were no reported injuries or adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-SERIES
Common Device Name
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
STRYKER-COMMUNICATIONS
1410 lakeside parkway #100
flower mound TX 75028
MDR Report Key11024320
Report Number0002031963-2020-00010
Device Sequence Number3562565
Product Code FQO
UDI-Device Identifier07613327400502
UDI-Public07613327400502
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number0682400550
Device Catalogue Number0682400550
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/23/2020
Supplement Date Received by Manufacturer11/23/2020
Initial Report FDA Received Date12/16/2020
Supplement Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-