| Model Number |
VENUL16080 |
| Medical Device Problem Codes |
Misfire (2532); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 08/2022).Device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the stent was allegedly failed to expand.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the sample was not returned to the manufacturer for evaluation.Therefore, the investigation is inconclusive for the reported failure to expand and partial deployment issues.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.In regards to a potential deployment failure the instructions for use state: 'predilation of chronic lesions with a balloon dilatation catheter is recommended.' and 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' in addition, the instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H11: d4 (product catalog no and corporate lot no), h6 (result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the stent allegedly failed to expand.The procedure was completed using another device.There was no reported patient injury.
|
| |
|