Model Number UNKNOWN
Device Problem Microbial Contamination of Device (2303)
Patient Problem Bacterial Infection (1735)
Event Type  Injury  
Manufacturer Narrative
Patient information was not provided. Serial number is unknown. This information will be provided in a supplemental report if made available. As the serial number is unknown, the device manufacture date could not be determined. This information will be provided in a supplemental report if made available. Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices. The z number is z-2076/2081-2015. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Through follow-up communication livanova learned that the device was cleaned as per the instruction for use and placed inside the operating theatre during use with the fan directed opposite to the patient at an estimated distance of 3. 5 meters from the surgery field. However, the serial number of the device used during surgery remains unknown. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received medwatch report mw (b)(4) stating that a (b)(6) years old male patient undergoing cardiac surgery in (b)(6) 2018 and an heater-cooler 3t system was used. Three (3) months later the patient undergone another surgery where cultures were taken and one month later he resulted positive to m. Chimaera.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
MDR Report Key11027708
MDR Text Key222064563
Report Number1718850-2020-01220
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberUNKNOWN
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/23/2020
Event Location No Information
Date Manufacturer Received11/23/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

Patient Treatment Data
Date Received: 12/17/2020 Patient Sequence Number: 1
-
-