| Model Number |
PROXIDIAGNOST N90 |
| Medical Device Problem Code |
Radiation Output Problem (4026)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Int.Ref.(b)(4).The investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
|
| |
|
Event or Problem Description
|
|
During annual physics testing, the stabilized dose is high during continuous fluoroscopy and pulsed fluoroscopy for 20 and 30 frames per second.
|
| |
|
Additional Manufacturer Narrative
|
|
Int.Ref.424305 the stationary nearby-controlled system supports general radiography and fluoroscopy imaging.The system has eleva workspot with eleva software integrated generator control, hand switch, keyboard, mouse, touch screen, and pc.There are various examination parameter settings available in eleva workspot.The investigation concluded that there was a software anomaly found in eleva software for dose calculation, which results in incorrect dose feedback causing the generator to regulate to (and stabilize at) a higher-than expected image dose.The issue started a capa investigation which led to a recall (fco 70600104).Philips field service engineer installed the fco.After this the risk is acceptable and the system is working as specified.The issue is further monitored and trended.Correction: h6 result and conclusion code.
|
| |
|