MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Overheating of Device (1437); Inappropriate/Inadequate Shock/Stimulation (1574); Connection Problem (2900)
|
Patient Problems
Burning Sensation (2146); Electric Shock (2554)
|
Event Date 12/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.It was reported that the rtm tried to "barbeque" them last night.Patient clarified the cord right where it went into the paddle of rtm got hot/shocked.Patient said it "bit the hell out of me".The cord looked like it pulled loose there where it went into the paddle.
|
|
Manufacturer Narrative
|
H6: recharger was returned for analysis and analysis found recharge antenna frayed - cable insulation was peeling away at donut end.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Additional information was received from a patient.It was reported that the patient replaced the cord.
|
|
Search Alerts/Recalls
|
|
|