|
Model Number G407371 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes, along with investigation results, will be provided in a subsequent submission.
|
|
Event Description
|
During the procedure, following insertion into the patient, air was aspirated from the hemostasis valve.The introducer was replaced, and the procedure was completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
One 8f swartz braided introducer sheath was received for evaluation.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak is consistent with damage during use.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
|
Search Alerts/Recalls
|
|
|