| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2020-04143.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial total knee arthroplasty, a scrub tech accidentally threw out the locking screw for a tibial insert.A screw was retrieved from a thinner tibial insert and was attempted to be used, but it would not lock the tibial insert into the tibial tray.The procedure was completed with a screw from another tibial insert of the same size.Attempts to obtain additional information have been made; however, no more is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The tibial plate item #00598004701, lot #j6811051 was not returned for evaluation as it remains implanted.Medical records were not provided.However, the information provided indicates that the decision to use the screw from a non-compatible product was an error, before a screw of the correct size was finally used.Device history record (dhr) was reviewed and no discrepancies were found.Review of the products determined that no failure was found as the products are within specification(s).No further investigation or action is required after review.The wrong screw (too short) was selected to assemble the articular surface to the tibial implant.The root cause of the reported issue is attributed to a user error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|