Model Number 72202618
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that, during an arthroscopic procedure, at the time of inserting the twinfix ultra 5.5 implant, it detached from the handle in which it came and bent when trying to re-pass it.The substitutions that came with it broke and the entire implant was detached, leaving it useless.In addition, one of the sutures of the needles disinserted the thread.The pieces of this implant were removed and another twinfix was used to successfully complete the surgery.No significant delay and no damage caused.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection revealed that the device was returned with one anchor and 4 needle sutures.The anchor had signs of wear from pliers or another grasping device.The sutures were cut and frayed in various locations, with debris present in the locations of the frays.The insertor of the device was significantly bent with some scratches present.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A clinical investigation found there were no patient injuries reported and no apparent patient impact based on the details provided.Therefore, no further medical assessment is warranted.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force or torque, off-axis insertion, or improper preparation of the insertion site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TF ULTRA 2 UBRAID AND NDLS 5.5MM TI
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11030020
Report Number1219602-2020-02077
Device Sequence Number1046954
Product Code MBI
UDI-Device Identifier03596010656131
UDI-Public03596010656131
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/02/2025
Device Model Number72202618
Device Catalogue Number72202618
Device Lot Number2049812
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/24/2020
Supplement Date Received by Manufacturer04/12/2021
Initial Report FDA Received Date12/17/2020
Supplement Report FDA Received Date04/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-