MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problem Overheating of Device (1437)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that per evaluation of the arctic sun device, i/o circuit card had the smell of burnt component. However, there were no physical signs on the circuit board. Later, the test i/o circuit card was installed into the device. It was also reported that the pressure transducer was reading 1. 0 psi at ambient and a test pressure transducer was installed to correct the reading to 0. 0 psi at ambient. It was further reported that a test circulation pump had to be installed to fill the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11030675
MDR Text Key222282182
Report Number1018233-2020-21773
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/27/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 12/17/2020 Patient Sequence Number: 1
-
-