• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER Back to Search Results
Model Number 407451
Device Problem Gas/Air Leak (2946)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2020
Event Type  malfunction  
Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the procedure, an air leak occurred.While placing the sheath into the patient, prior to catheter insertion, air was aspirated into the syringe connecting to the extension port of the sheath.The device was replaced and the procedure was completed with no adverse patient consequences.
 
Manufacturer Narrative
One 8.5f swartz braided introducer sheath was received for evaluation.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 63 CM LENGTH, 8.5 F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key11032209
MDR Text Key222557487
Report Number3005334138-2020-00603
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205757
UDI-Public05414734205757
Combination Product (y/n)N
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model Number407451
Device Catalogue Number407451
Device Lot Number7476418
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-