ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 407451 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results, method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During the procedure, an air leak occurred.While placing the sheath into the patient, prior to catheter insertion, air was aspirated into the syringe connecting to the extension port of the sheath.The device was replaced and the procedure was completed with no adverse patient consequences.
|
|
Manufacturer Narrative
|
One 8.5f swartz braided introducer sheath was received for evaluation.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak is consistent with damage during use.
|
|