| Model Number |
482619580 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while putting the set screw on, the xia serrato polyaxial screw tulip head disengaged intra-operatively.The procedure was completed successfully with an unknown surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device and complaint history records were reviewed for this lot, no relevant manufauring issues or similar complaints were identified.Initial report stated that the screw head came off screw while putting set screw on.Follow up confirmed that this screw had been implanted for 2-3 years.While tightening a set screw onto the tulip, the tulip head disengaged.Root cause could not be conclusively determined as the device was not returned.Most likely root cause is due to excessive force or overtightening of blocker.Misalignment of blocker before tightening may have also contributed to the event.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that while putting the set screw on, the xia serrato polyaxial screw tulip head disengaged intra-operatively.All the fractured pieces were removed from the surgical site and the procedure was completed successfully with a 30 minute delay.No adverse consequences or medical intervention were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that while putting the set screw on, the xia serrato polyaxial screw tulip head disengaged intra-operatively.All the fractured pieces were removed from the surgical site and the procedure was completed successfully with a 30 minute delay.No adverse consequences or medical intervention were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated b5 with the additional surgical delay information received.
|
| |
|