| Catalog Number |
0375542000 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: a red stripe was discovered at the distal end of the shaver blade the failure alleged in the complaint record was not confirmed during the product investigation.Based on the alleged claim the probable root cause/s could be a small amount of plastic (flash) was created during the inductive process or an insertion of the inner cutter accidentally scraped the housing hub and remained attached to the cutter teeth.The improper cleaning process, qc incoming, and in-process inspections.An nc was opened to address the issue.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|