| Model Number |
305313 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Underdose (2542)
|
| Date of Event |
11/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Investigation conclusion: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Root cause description: no root cause can be determined as no samples were received.Rationale: since no samples displaying the reported condition were received corrective actions are not necessary.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle integra 22x1-1/2 rb tw leaked during a "long acting deep intramuscular injection", causing an underdose of medication.The following information was provided by the initial reporter: "t2 failures in 7 days, leakage of oil based long acting deep intramuscular injection.This caused reduced medication dosage administration.".
|
| |
|
Search Alerts/Recalls
|