| Catalog Number |
LC210 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Failure to Form Staple (2579); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4); batch #: unk.Date of event is 2020.Event day and event month were not reported.The serial number was not provided, therefore, the manufacturing record evaluation could not be performed.Attempts have been made to retrieve the device.To date, the device has not been returned.If the device, or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the clip applier didn't apply proper clips.The jaws were spoiled.User reported the clips are not broken, the clip applier didn't hold the clips.Clip applier jaws were broken so they didn't hold the clips and because the clip applier jaws were broken, it didn't make the clips hold the vein, or tissue.It is unknown how procedure was completed.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/13/2021.H6: component code: mechanical (g04).Investigation summary: the analysis results confirmed that the lc210 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.No conclusion could be reached as to what may have caused the found condition of jaws.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.The serial number history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment. the certificate records are accessible through external manufacturing. .
|
| |
|