| Model Number |
RONYX20015UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a mildly tortuous, moderately calcified lesion exhibiting 70% stenosis located in the distal right coronary artery.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during removal following a failed delivery.It was stated that the stent would not advance into the rca and the stent was pulled back.It was stated that the stent got stripped of the balloon on removal and dislodged into the paravertebral lumbar artery l3.The dislodged stent was not removed from the patient and after a whole body ct it was decided to not remove the stent and follow up with it.The patient was reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: there was no difficulties noted when removing the protective sheath.The stent was inspected post removal of the protective sheath with no issues.The inflation device remained on neutral pressure during delivery of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: image evaluation summary: one still image was provided for review.The image provided shows an mri scan of the torso.Pen is pointed at an area of the abdomen however it is difficult to identify the stent.Image does not confirm stent dislodgement during removal.Device evaluation summary: device returned for evaluation.The stent was not present on the balloon and did not return for analysis.Crimp impressions were visible on the exposed balloon surface.The balloon folds remained intact.The inner lumen patency was verified with a 0.015 inch mandrel.Deformation was evident to the distal tip.No other deformation was noted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|