WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.0MM CRUCIFORM SCREWDRIVER BLADE WITH HOLDING SLEEVE Back to Search Results
Model Number 314.48.96
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a synthes employee.The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during an inspection at the warehouse on an unknown date, the cruciform screwdriver blade with holding sleeve tip is broken.There was no patient involvement.This report involves 1 1.0mm cruciform screwdriver blade with holding sleeve.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h3, h4, h6: part 314.48.96, lot 4516676: release to warehouse date: december 03, 2002.Manufactured by synthes brandywine.No non-conformance reports (ncr's) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.H3, h6: a product investigation was completed: upon visual inspection it was noticed that the tip was broken.No other defects were identified on the device.No dimensional analysis was performed due to confirmed post manufacturing damage identified.The relevant drawings were reviewed.The complaint condition is confirmed.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: corrected data: g1: physical manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.0MM CRUCIFORM SCREWDRIVER BLADE WITH HOLDING SLEEVE
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11036938
Report Number2939274-2020-05645
Device Sequence Number2692015
Product Code HXX
UDI-Device Identifier10886982188946
UDI-Public(01)10886982188946
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number314.48.96
Device Catalogue Number314.48.96
Device Lot Number4516676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Initial Date Received by Manufacturer 11/24/2020
Supplement Date Received by Manufacturer01/07/2021
Initial Report FDA Received Date12/18/2020
Supplement Report FDA Received Date01/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-