| Model Number |
9-ASD-019 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes, along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 19mm amplatzer septal occluder was selected for implant.During procedure, the left atrial disc formed a cobra shape, and the device had difficulty being re-sheathed to be recaptured.The physician successfully recaptured the device, and it was removed from the patient without further incident.Then, the patient was referred for surgery for surgical closure.Of note, it was reported that the patient experienced a sudden cardiac arrest during the second surgery and on (b)(6) 2020, the patient passed away.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of deformity during deployment not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|
| |
|