DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/02/2020
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that cable of the device was ripped off.Per service reports, this complaint can be confirmed.It was found during evaluation that the motor was corroded.Further, the motor cable and keyboard resistance value were out of tolerance limits.The motor, motor cable and keyboard were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Fluid ingress into the device and contact with the motor is responsible for the rusty motor.With the available information, we cannot determine the root cause of the defective motor cable and defective keyboard.The corroded motor and defective motor cable and defective keyboard would have caused the customer to experience the reported problem.This serialized device ((b)(4)) was manufactured prior to the current manufacturer, therefore a dhr review cannot be performed since the original manufacturer no longer exists and therefore the manufacturer can no longer make any process correction or corrective actions if needed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: a review of lot/batch history for each legacy fms product complaint received by mitek chu is not recommended since legacy manufacturing no longer exists, and it is no longer possible to make process corrections or corrective actions.This device (serial number: (b)(4)) was produced prior to the closure of the fms facility in nice, france and therefore will not have a dhr review preformed.Please see signed legacy fms batch review.
 
Event Description
It was reported by the affiliate in (b)(6) that the cable on the micro tornado hp w handcontrol device was ripped off.During in-house engineering evaluation, it was determined that the motor was rusty.There was no procedure involved.There were no adverse patient consequences.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11039258
MDR Text Key222817429
Report Number1221934-2020-03946
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2020
Date Manufacturer Received12/02/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-