ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SSL-380
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "left side rupture".Device has been explanted.
 
Event or Problem Description
Healthcare professional reported "left side rupture".Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified one curved opening and fold creases.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identify one opening crease sharp, stress marks and wear abrasion.Based on the device analysis the final assessment is: one opening crease sharp in the side radius assessed as fold flaw opening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key11039735
Report Number9617229-2020-20908
Device Sequence Number1023399
Product Code FTR
UDI-Device Identifier10888628032576
UDI-Public10888628032576
Combination Product (Y/N)N
Initial Reporter StateOR
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/15/2021
Device Model NumberSSL-380
Device Catalogue NumberSSL-380
Device Lot Number3001234
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Initial Date Received by Manufacturer 11/20/2020
Supplement Date Received by Manufacturer12/22/2020
Initial Report FDA Received Date12/18/2020
Supplement Report FDA Received Date01/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-