Model Number E053
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 10/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.(b)(4).
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was part of a system revision due to infection.The patient was treated with intravenous antibiotics.There were no additional adverse patient effects reported.The icd was explanted.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Patient code 3191 captures the reportable event of surgery and the additional intervention performed by the use of intravenous antibiotics.Upon receipt at our post market quality assurance laboratory product investigation indicates that the that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.Infection is a known medical risk when the skin barrier is breached.Analysis of the returned product was not able to provide relevant information for infection-related allegations.Additional information is pertinent as this provides significant impact that changed the overall understanding of the event.Additional information does not impact the current reporting decision.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was part of a system revision due to infection.The patient was treated with intravenous antibiotics.There were no additional adverse patient effects reported.The icd was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA ICD
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key11040274
Report Number2124215-2020-26794
Device Sequence Number1265616
Product Code LWP
UDI-Device Identifier00802526480706
UDI-Public00802526480706
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S235
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/06/2015
Device Model NumberE053
Device Catalogue NumberE053
Device Lot Number101272
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer01/20/2021
Initial Report FDA Received Date12/18/2020
Supplement Report FDA Received Date01/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age31 YR
-
-