ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number MM-410360
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 10/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified, the weight the device within specification, crease fold, deformation, wear abrasion, and white particles on the shell.After autoclave cycle were observed: cloudy color in the gel and voids.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key11040513
Report Number9617229-2020-20858
Device Sequence Number13708935
Product Code FTR
UDI-Device Identifier10888628003644
UDI-Public10888628003644
Combination Product (Y/N)N
Initial Reporter StateNC
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/21/2019
Device Model NumberMM-410360
Device Catalogue NumberMM-410360
Device Lot Number2666292
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Initial Date Received by Manufacturer 11/20/2020
Supplement Date Received by Manufacturer12/21/2020
Initial Report FDA Received Date12/18/2020
Supplement Report FDA Received Date01/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
-
-