| Model Number |
7K78-30 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Patient identifier was truncated; full sid is (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer generated a falsely elevated architect total b-hcg result for one patient.The following information was provided: sid (b)(6) initial result: 69.43 miu/ml, retest 94.23, 90.7, 83.49, 90.23 and 90.3 miu/ml (positive) retest on roche platform: 92.9 miu/ml (positive).New draw (sid unknown), retest on architect: <1.2 miu/ml, retest <1.2 miu/ml and <1.2 miu/ml (negative).The patient is a (b)(6) year old female with abnormal uterine bleeding/severe anemia.The patient had undergone a blood transfusion prior to the initial blood draw.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets was performed for reagent lot number 11237ui00.The ticket search determined that there is elevated complaint activity for the reagent lot.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.A retained reagent kit of lot number 11237ui00 was tested.All specifications were met during completion of the protocol, demonstrating that the lot is performing acceptably.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the architect total b-hcg reagent lot 11237ui00 was identified.
|
| |
|
Search Alerts/Recalls
|