SMITH & NEPHEW, INC. GNS II RESURF PAT 35MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420578
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, during a tkr washing, at the time of opening the patient, there was a large amounts of pus present, so the tkr implants were removed due to infection and cement spacer was inserted.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical / medical investigation concluded that, this case reports that a tkr washout procedure was planned, however when the patient was opened up, there was large amounts of pus present.Therefore, due to infection, all of the tkr implants were removed and an antibiotic spacer was inserted.Per email communication, the requested surgical reports and x-rays will not be provided.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, no further clinical assessment is warranted.Should additional clinically relevant documentation become available, the clinical/medical task may be re-opened.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical / medical investigation concluded that, one undated x-ray photo was provided for review; however, it does not contribute to the root cause of the reported infection.Based on the limited information provided the clinical root cause for the reported infection could not be determined.There is no evidence to support that the smith and nephew device contributed to the reported infection.The future impact to the patient beyond the revision cannot be determined.Should additional information become available this issue can be re-elevated.No further clinical assessment is warranted at this time.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II RESURF PAT 35MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11042020
Report Number1020279-2020-07594
Device Sequence Number2692010
Product Code JWH
UDI-Device Identifier03596010208316
UDI-Public03596010208316
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420578
Device Catalogue Number71420578
Device Lot Number18KM22725
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer07/29/2021
08/31/2021
Initial Report FDA Received Date12/18/2020
Supplement Report FDA Received Date07/29/2021
09/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-