| Model Number |
FRED3021-PMA |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Speech Disorder (4415); Thrombosis/Thrombus (4440)
|
| Date of Event |
11/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformance's associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for analysis.Additionally, procedural images were not provided; therefore, the alleged event cannot be confirmed.The instructions for use identifies stent thrombosis and thromboembolic event as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that 12 hours after implantation of the fred in the supraclinoid segment of the ica, thrombus was identified in the artery.The patient experienced dysarthria.Reopro and heparin were administered to the patient.There was no reported malfunction of the fred.
|
| |
|
Search Alerts/Recalls
|