|
Reported event: an event regarding wear of a triathlon insert was reported.The event was confirmed by inspection of received image of device.Method & results: - product evaluation and results: the device was not returned.Visual inspection of the received image of the device noted the insert to be worn near one of the edges.Dimensional, functional and material analysis were not performed as the device was not returned.- clinician review: the available medical records were provided to the consulting clinician for a review which noted, " the event was confirmed, there was a revision performed for polyethylene wear (note: images and intraoperative photos were unlabeled so side cannot be confirmed).The root cause cannot be defined without additional medical records, x-rays, intraoperative findings, and perhaps examination of the explant.It appeared that the polyethylene had worn and delaminated at the posteromedial corner of the tibia.This was possibly related to instability of the joint and malrotation especially at this early timepoint." - device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.- complaint history review: there have been no other similar events for the reported lot.Conclusions: it was reported that the patient was revised due to insert wear.Visual inspection of the received image of the device noted the insert to be worn near one of the edges.The available medical records were provided to the consulting clinician for a review which noted that the event was confirmed, there was a revision performed for polyethylene wear (note: images and intraoperative photos were unlabeled so side cannot be confirmed).The root cause cannot be defined without additional medical records, x-rays, intraoperative findings, and perhaps examination of the explant.It appeared that the polyethylene had worn and delaminated at the posteromedial corner of the tibia.This was possibly related to instability of the joint and malrotation especially at this early timepoint.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pre and post operative x-rays, operative reports as well as patient history and follow up notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If device is returned or additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|