WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-1-JUG-CELECT-PT |
Device Problems
Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Detachment of Device or Device Component (2907); Unintended Movement (3026); Insufficient Information (3190)
|
Patient Problems
Internal Organ Perforation (1987); Thrombosis (2100); Perforation of Vessels (2135); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook celect platinum filter.Reporter occupation: non-healthcare professional.Pma/510(k) k171712.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect platinum filter on (b)(6) 2018.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
|
|
Manufacturer Narrative
|
Additional information: investigation the following allegations have been investigated: vena cava (vc)/organ perforation, thrombosis, tilt, pain, swelling, discomfort, physical limitations.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain, swelling, discomfort, and physical limitations are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Twenty devices were manufactured in this lot.To date, one unrelated complaint has been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2018 via the right internal jugular vein due to deep vein thrombosis.Patient is alleging tilt, vena cava perforation, organ perforation, and caval thrombosis.Patient further alleges pain in midsection, back, upper torso and neck, swelling, discomfort during movement, sharp pains in mid and upper torso and upper right quadrant, right arm and shoulder which is extreme during activity as well as limitations in upper body activity and all daily activities.Report from computerized tomography (ct): "there is an ivc filter from the mid l to level to the l3-4 disc level.The major and minor prongs appear intact.No missing prongs are identified.The filter is slightly tilted such that the proximal tip abuts the posterior left wall of the ivc.The major prongs each penetrate the ivc wall with least penetration posteriorly.The medial prong abuts the right lateral wall of the aorta.Report ct: "ivc filter in place with either ivc thrombus or inhomogeneous contrast opacification of the ivc inferior to the filter.".
|
|
Manufacturer Narrative
|
Investigation the following allegations have been investigated: fracture, migration.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
27dec2018, per a report from computed tomography 2; ¿impressions: the filter is tilted to the posterior left with its proximal cone against the ivc wall.Only 7 filter arms are seen.There should be 8 arms.One arm is missing which represents a fracture with migration of the arm off the image set.The anterior strut penetrates 8 mm through the ivc wall.The left strut penetrates 5 mm through the ivc wall.The posterior strut penetrates 0 mm through the ivc wall.The right strut penetrates 4 mm through the ivc wall.¿.
|
|
Search Alerts/Recalls
|
|
|