| Model Number |
VENUL14060 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending; however, images were provided for review.The investigation of the reported event is currently underway.Expiry date (09/2022); device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement, the device allegedly failed to expand.It was further reported that after manipulation, the stent finally deployed after five minutes.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned to the manufacturer for evaluation.An image was provided for review.The condition of the returned delivery system confirmed the alleged expansion issue.The silicone brake of the inner catheter which is under the stent was found broken and shifted to distal which indicated that the stent adhered to the brake leading to breakaway and shifting upon removal.Two images were provided for two complaints, but it was not known which image belonged to this complaint.The condition of the sample leads to a confirmed result for expansion issue.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of the relevant for this product was conducted.The instruction for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 09/2022), g3.H11: h6 (method, result).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement, the device allegedly failed to expand.It was further reported that after manipulation, the stent finally deployed after five minutes.There was no reported patient injury.
|
| |
|