Model Number 78-30020
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
 
Event or Problem Description
It was reported by the company representative that the patient developed an infection post operative, and a revision surgery was performed to remove the product.
 
Additional Manufacturer Narrative
The reported event could not be confirmed.Additional information (e.G.Initial implantation date, date the infection was discovered, and laboratory results) were requested regarding the reported event.Further, the sales rep was asked to fill-out the ¿infection complaints ¿ checklist customer¿ (cmfqf (b)(6) ).No further information could be provided except the initial implantation date, (b)(6) 2020.The ¿infection complaints ¿ checklist investigator¿ (cmfqf (b)(6) ) was completed by the manufacturer for the affected device.For infection complaints, expanded investigations were performed.This is to assure that neither the reported device nor all related manufacturing (e.G.Environmental monitoring, sterilization, validations tests) process steps could have caused the event.Indications for any device or manufacturing problems were not found.The reported device is delivered non-sterile.The related instructions for use (90-02023) include the appropriate cleaning, disinfection, and sterilization instructions for the device.Infection is a known adverse event related to this procedure.In general, infections, foreign body reactions, and inflammations following the insertion of implants are known adverse events associated with these procedures.They are documented in the instructions for use and have been considered in the risk evaluation of the relevant stryker product.This type of adverse event may rather be clinically related than implant related.Based on statistical evaluation, there is no indication for an incorrectly working product or any systematic design, material or manufacturing related issue.Therefore, no corrective and / or preventive actions are deemed necessary at this time.The complaint is added to the complaint trend.H3 other text : device not available for return.
 
Event or Problem Description
It was reported by the company representative that the patient developed an infection post operative, and a revision surgery was performed to remove the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOMIZED MAND. RECON PLATE20, HEMI
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
MDR Report Key11045851
Report Number0008010177-2020-00066
Device Sequence Number11018834
Product Code JEY
UDI-Device Identifier07613252599647
UDI-Public07613252599647
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K132519
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number78-30020
Device Catalogue Number78-30020
Device Lot Number2007251001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/03/2020
Supplement Date Received by Manufacturer12/03/2020
Initial Report FDA Received Date12/21/2020
Supplement Report FDA Received Date04/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-