| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Codes |
No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical product: catalog #: 407398, comp primary stem inserter, lot # 287480; catalog #: unknown, reverse tray, lot # unknown.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a comprehensive reverse total shoulder procedure a humeral stem could not be attached to the inserter.The stem was inserted by hand, and then the humeral tray would not engage the stem.The stem was then removed.Additional broaching was required and a larger size stem was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Updated: b1, b4, b5, g3, h1, h2, h3, h6, h10.Complaint was confirmed by visual examination of the returned product which identified that there is extensive damage to the stem in terms of scratches and taper damage.The inserter is missing the spring that holds tension to the rod that holds the stem.The set screw that holds the spring into place in the inserter was loose and does not show any signs of the epoxy anymore.Inside the inserter where the set screw is installed does not show any signs of epoxy as well.The inserter was manufactured in may of 2018.The stem and the inserter do mate as intended but the trigger had to be pushed towards the stem since the spring is missing.Device history record was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue is attributed to user error as the spring was missing from the inserter causing it to not function.This should have been identified prior to the procedure and/or during reprocessing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|