• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY OVERSTITCH ENDOSCOPIC SUTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY OVERSTITCH ENDOSCOPIC SUTURING SYSTEM Back to Search Results
Model Number CNH-G01-000
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528); Premature Separation (4045)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was returned to the apollo device analysis laboratory on 11/dec/2020.Analysis of the device is ongoing.A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential events of "helix-did not turn properly" and "helix- could not be removed from tissue" as follows: "warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures." "caution: if resistance is encountered when advancing the helix through the working channel of the endoscope, reduce the endoscope angulation until the device passes smoothly, and ensure that the secondary working channel is not obstructed on the endoscope." "caution: do not depress helix handle button while advancing helix through endoscope." "warnings: do not use a device where the integrity of the sterile packaging has been compromised or if the device appears damaged." warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Ensure that there is sufficient space for the needle to open.Warning: do not introduce the device with the needle body in its open position.
 
Event or Problem Description
The tissue helix got stuck in the tissue.The patient was treated successfully.
 
Additional Manufacturer Narrative
Supplemental #1 medwatch submitted to the fda on 05/feb/2021.A discrepancy in our logging system was discovered on 02/feb/2021 in which the wrong complaint had been attached to the record.Corrections have been made and this complaint is no longer reportable as it was previously reported as mdr #: 3006722112-2020-00113.Corrections: b1, b3, b5, d4, h4, h6 additional info: h10.
 
Event or Problem Description
A cinch came loose from the suture prematurely, through which the seam couldn't be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERSTITCH ENDOSCOPIC SUTURING SYSTEM
Common Device Name
SUTURING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key11046350
Report Number3006722112-2020-00129
Device Sequence Number1265450
Product Code OCW
UDI-Device Identifier10811955020688
UDI-Public10811955020688
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K081853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2022
Device Model NumberCNH-G01-000
Device Catalogue NumberCNH-G01-000
Device Lot NumberAF03701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer12/11/2020
Initial Report FDA Received Date12/21/2020
Supplement Report FDA Received Date02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient Weight92
-
-