Other, other text: device evaluation: three device samples were returned for analysis in used condition without their original packaging.An air cuff symmetry test was performed and the investigation found that when the samples were inflated with air, the cuff inflated completely.When measuring the cuffs, the percentage for the symmetry was for sample 1: 50%, sample 2: 50%, sample 3: 44.4%.These results are over 33.3% that is the minimum acceptable.Based on the test the units were within specification.Based on the investigation, the complaint allegation was not confirmed.
|