| Model Number |
TENODESIS SCREW, PEEK |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a thumb ucl reconstruction the screw could not be detached from the handle.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a different device from another manufacturer.The surgeon has switched to a different technique to finish the procedure.Update 14-dec-2020: further information was provided that no additional incision was necessary due to the technique change.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the screws were found to be tight but came off without being damaged.The driver tips were found to meet specifications but the screws internal hex features could not be measured due to the deformation observed.The cause is undetermined.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the screws were found to be tight but came off without being damaged.The driver tips were found to meet specifications but the screws internal hex features could not be measured due to the deformation observed.The cause is undetermined.
|
| |
|