| Model Number |
1488010000 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: light source turns off mid case.Probable root cause: cables; connectors; digital board; power supply; filter / fuse; ac inlet board; main board; transition board; fiber board; intermittence between transition board and ch connecto; software; scope; light source; camera head; coupler; 1488 & 1588 extension cable; intermittence while using rf devices; problem toggling light source or from camera head; connected monitors; leaking; use error; poor grounding; electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge; manufacturing/ service nonconformity.Manufacture date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|