| Model Number |
100/897/090CZ |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received indicating that a smiths medical uni perc size 9 adjustable flange tracheostomy was inserted in patient on (b)(6) 2020.On (b)(6) 2020 the connector (white section that ventilator circuit attaches to) separated from the tube.As a result tracheostomy change required.There were no reported adverse events.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: one portex tube was returned for analysis.Under visual inspection, it was noted that the white connector is separated from tracheostomy tube as described by customer.There was not any other defect found on returned sample.Connectors are manually bonded with tracheostomy tube as per the manufacturing procedure by the help of the solvent.Within this manufacturing process there are also implemented several quality checks which contains visual and functional inspection.The returned sample was inspected under magnification and presence of solvent was confirmed.There are solvent spots on the connector which indicates that solvent process happened.Complaint history was checked and no trend of confirmed customer complaints was identified.Therefore this issue is considered to be isolated incident with unknown root cause.
|
| |
|