DEPUY ORTHOPAEDICS INC US SIG PS CEM FEM W/LUGS RT SZ4N; SIGMA KNEE PRIMARY : KNEE FEMORAL Back to Search Results
Model Number 1960-79-500
Medical Device Problem Code Off-Label Use (1494)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient's knee was revised to address loosening of the tibial component at the cement to implant interface.Cement manufacturer unknown.Surgeon noted construct was off label: 4 rp ps femur with 2.5 mbt tibia was explanted.Implant debonded from cement mantel.No surgical delay.Doi: (b)(6) 2012, dor: (b)(6) 2020, right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.H10 additional narrative: added: d7a, d9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG PS CEM FEM W/LUGS RT SZ4N
Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11049420
Report Number1818910-2020-27378
Device Sequence Number11076782
Product Code JWH
UDI-Device Identifier10603295075530
UDI-Public10603295075530
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K073529
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1960-79-500
Device Catalogue Number196079500
Device Lot NumberE68F34
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/07/2020
Supplement Date Received by Manufacturer01/19/2021
Initial Report FDA Received Date12/21/2020
Supplement Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT CEM KEEL TIB TRAY SZ2.5; PFC SIGMARP STB TB IN 4 10.0; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient Weight83
-
-