Model Number 60-02-50 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp was used as back up during a procedure and displayed an error code associated to the eprom when turned on.Reportedly the procedure could be completed with no issues.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: through follow-up communication livanova learned that the reported error code could not be reproduced.However, the processor board where the eprom is located was replaced as potentially involved component.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The most likely root cause of the reported event is a failure of the eprom on the processor board.
|
|